BIOSCOPE-C
Range of application:
- Emergency service
- Out-patient visits
- Pacemaker check-up
Special features
The BIOSCOPE is easy to use. By integrated 3-point electrodes the ECG can be detected on the patient’s chest. By using the cable adapter it is possible to use a patient cable.
There is a memory capacity of up to 11 minutes and the stored information can be retrieved by displaying it on the screen of the BIOSCOPE.
By means of an optional analog adapter (BIOSCOPE-AI) or a digital adapter (BIOSCOPE-DI) respectively the stored ECG can be transmitted to an external ECG recorder or to a PC. In the later case the software BIOTREND features extended analysis functions. The BIOSCOPE-C is also used in our Anaesthesia Monitor BIO-3.
Optional accessories: Cable adapter and patient cable, Analog interface for recording the stored ECG’s, Digital interface connecting the PC, Software BIOTREND

Please send an email to sales@recomedical.com to ask for pricing.
Operating instructions and software updates you will find here:
| Lightweight, portable and battery powered n | 6LR61, 9V |
|---|---|
| Вimensions | 52 x 70 x 40 mm |
| Weight | 350g |
| LCD screen for ECG and heart rate | |
| Memory capacity | up to 11 minutes |
| Nennfrequenz | 2.450 MHz ±50 MHz |
| Three recording speeds ( | 25, 50 und 100 mm/sec |
| Automatic adjustment of amplification | 0,25 mV … 8 mV |
| Display of ST-depression | |
| Detection and display of pacemaker pulses | |
| ECG cable for external (e.g. adhesive) electrodes | |
| Interface to connect standard ECG recorder (optional) | |
| Interface to PC, incl. software BIOTREND for comprehensive ECG analysis (optional) | |
| Bandwidth | 0.5 … 40 Hz (3 dB) |
| Input impedance | 2 MOhm |
| Class IIa MDD 93/ 42/ EEC, IEC 60601-1, IEC 60601-2-27 (defi protected by using cable adpter CA-2) | |
| CE 0123 |